BYC · ConfidentialUnfiled inventionModel-only · 26–27 Nov 2026
Power feasibility · study report
Acoustic Contact Lens
Can a battery in smart glasses power the lens across the ear? Evaluated in COMSOL — a coil and planar-array primary transducer, and a secondary receiver coil with a Balanced Armature for now. Model-only — nothing measured yet.
videoTemple coil feeding the ear — winding geometry from COMSOL.videoIn-ear assembly — coil, passive ring and the lens in the canal.
COMSOL renders, not photographs — no hardware exists yet.
Short answer
On paper, yes — the current study supports the feasibility.
One smart-glasses battery runs one lens for a full day and more: 9–28 h on the figures the model reports, 43–94 h if one bookkeeping step is corrected. Hearing glasses on sale today run 8–10 h.
1.2 mW
received by the lens for up to ~105 dB at the eardrum
6–58 mW
drawn from the battery, per ear — all designs, both ways of counting
0.6 Wh
energy in one 160 mAh cell (first Ray-Ban Meta)
Per ear: each arm of the glasses has its own battery and coil. dB is the sound level at the eardrum.
Two counts throughout. The model reports say the lens needs 7.6 mW, but ~65% of that is lost in the transmitting coil and the lens itself took only ~1.15 mW — so the reports look ~6× high. Corrected and as-reported are shown side by side; a circuit-model re-run will settle it.
How the power travels
Power and sound cross 3 cm of air
The lens has no battery. The glasses send power and sound together as a magnetic field at 6.78 MHz.
Battery
in the temple
Amplifier
makes the RF drive
Temple coil
3 turns in the arm
· · · 3 cm air · 6.78 MHz · · ·
Copper ring
at the ear · passive · optional, doubles the share that arrives
Acoustic Contact Lens
coil · 1 diode · Balanced Armature
Eardrum
sound out
The device
A coil, one diode, a Balanced Armature
Secondary: a receiver coil with a Balanced Armature for now — the in-canal transducer, which can be swapped later.
COMSOL geometry — temple primary, the ~3 cm gap and the node in the canal, on the head model.
Node cut through its axis — ferrite core, 0.26 mm oval foil, 2-turn receive coil (copper strip 1.35 × 0.10 mm), and the Balanced Armature parts.
Where the power goes
Almost none of it becomes sound
Temple + ring, corrected count: 2.9 mW from the battery for the link.
Amplifier loss0.87 mW
Lost crossing the gap (coils + ring)0.89 mW
Lost in the diode0.53 mW
Steady current through the Balanced Armature0.63 mW
The sound signal is just 0.04 mW at 96 dB and 0.27 mW at 105 dB. The lens has no battery and no chip — one diode does all the work, and a steady current flows through the Balanced Armature even in silence.
Biggest lever: block that steady current with a capacitor, or add a small chip — either could cut the link power several times over.
Headroom
Glasses batteries are big enough
HA button cell175 mAh
0.24 Wh
Small glasses cell100 mAh
0.39
Ray-Ban Meta G1160 mAh
0.62
Ray-Ban Meta G2210 mAh
0.81
Ray-Ban Display248 mAh
0.95
00.51.0 Wh
A whole hearing aid runs on 1.3–1.7 mW; a glasses cell holds 2.5–4× a button cell. Room for the link.
The real cost driver
Louder costs more
Temple + ring · 160 mAh · hearing mode. Every +10 dB ≈ 10× the link power.
correctedreported
Up to ~105 dBfirst target
89 h
25
Up to ~111 dB4× link power
36 h
7
Up to ~115 dB10× link power
16 h
3
0 h50100 h
105 dB suits mild loss. Target loudness is an open decision — set it with an audiologist.
Fit sensitivity
Where the glasses can sit
The temple primary moved as the frame slips (±2/±4 mm along and across, ±5° tilt), wound vs planar primary.
A downward slip can actually help; forward and upward slips cost the most link efficiency — up to about −30%.
Heat & exposure
Comfortable margins
>10×
margin on heat: link puts 2–39 mW into the temple coil; BYC's limit for a worn coil is 450 mW.
0.12 W/kg
peak absorbed by the head at 27 mW into the coil (limit 2 W/kg). The head takes ~2% of the coil's loss.
COMSOL field & local-SAR maps on the plane through the lens and the primary — wound primary (top) vs planar flex-PCB primary (bottom). Uniform block stands in for the head; not a safety assessment.
What could still break this
Nothing is measured. Coil quality, ferrite and the Balanced Armature case are assumed — real coils are usually worse.
The ring needs 20 mm at the ear; without it the link needs ~2× power.
Fit matters. Temple 5 mm up / 8 mm back ≈ −10 points of efficiency.
Amplifier assumed 70%. At 50%, the ring design gives 74 h, not 89 (corrected).
Loudness is rough. Speaker model failed 2 checks → read "105 dB" as ±several dB.
Mics + processor assumed 3 mW/ear; beam-forming may need more.
Near-term
Re-run the circuit model with the new lens → settles 1.25 vs 7.6 mW.
Bench test: wind the temple coil + 20 mm ring, measure arriving power vs model.
Kill the idle current (blocking cap or tiny chip) and model the gain.
Fix target loudness with an audiologist — it sets the power more than any coil detail.
Where this goes next
Ultrasonic MEMS
Evaluate as the in-canal transducer — the step beyond the Balanced Armature.
Array optimization
Optimize the coil / planar-array primary for higher link efficiency and steering.
Other acoustic transducers
The Balanced Armature can be re-evaluated against a planar acoustic transducer in the canal.
Where each input comes from · sources
Link efficiency 20–69% — COMSOL, run 2026-11-27_acl_temple_versions_v1.
Circuit power split (detector takes 1.15 mW of 7.58 mW) — COMSOL 2026-11-26_acl_eval_v1; diode data assumed.
Lens need 1.25 mW — inferred 1.15 × 1.09; to be confirmed.
Sound level — 96 dB @30% / 105 dB @80% modulation, 1 kHz; speaker checks G4, G5 failed.
Amplifier 70% (50% as a check) — a published 6.78 MHz implant link reports 52–68% end-to-end at 80 mW.